TY - JOUR
T1 - The accuracy of pulse oximetry in measuring oxygen saturation by levels of skin pigmentation
T2 - a systematic review and meta-analysis
AU - Shi, Chunhu
AU - Goodall, Mark
AU - Dumville, Jo
AU - Hill, James
AU - Norman, Gill
AU - Hamer, Oliver
AU - Clegg, Andrew
AU - Watkins, Caroline Leigh
AU - Georgiou, George
AU - Hodkinson, Alexander
AU - Lightbody, Catherine Elizabeth
AU - Dark, Paul
AU - Cullum, Nicky
N1 - Publisher Copyright:
© 2022, The Author(s).
PY - 2022/8/16
Y1 - 2022/8/16
N2 - Background: During the COVID-19 pandemic, there have been concerns regarding potential bias in pulse oximetry measurements for people with high levels of skin pigmentation. We systematically reviewed the effects of skin pigmentation on the accuracy of oxygen saturation measurement by pulse oximetry (SpO2) compared with the gold standard SaO2 measured by CO-oximetry. Methods: We searched Ovid MEDLINE, Ovid Embase, EBSCO CINAHL, ClinicalTrials.gov, and WHO International Clinical Trials Registry Platform (up to December 2021) for studies with SpO2–SaO2 comparisons and measuring the impact of skin pigmentation or ethnicity on pulse oximetry accuracy. We performed meta-analyses for mean bias (the primary outcome in this review) and its standard deviations (SDs) across studies included for each subgroup of skin pigmentation and ethnicity and used these pooled mean biases and SDs to calculate accuracy root-mean-square (Arms) and 95% limits of agreement. The review was registered with the Open Science Framework (https://osf.io/gm7ty). Results: We included 32 studies (6505 participants): 15 measured skin pigmentation and 22 referred to ethnicity. Compared with standard SaO2 measurement, pulse oximetry probably overestimates oxygen saturation in people with the high level of skin pigmentation (pooled mean bias 1.11%; 95% confidence interval 0.29 to 1.93%) and people described as Black/African American (1.52%; 0.95 to 2.09%) (moderate- and low-certainty evidence). The bias of pulse oximetry measurements for people with other levels of skin pigmentation or those from other ethnic groups is either more uncertain or suggests no overestimation. Whilst the extent of mean bias is small or negligible for all subgroups evaluated, the associated imprecision is unacceptably large (pooled SDs > 1%). When the extent of measurement bias and precision is considered jointly, pulse oximetry measurements for all the subgroups appear acceptably accurate (with Arms < 4%). Conclusions: Pulse oximetry may overestimate oxygen saturation in people with high levels of skin pigmentation and people whose ethnicity is reported as Black/African American, compared with SaO2. The extent of overestimation may be small in hospital settings but unknown in community settings. Review protocol registration: https://osf.io/gm7ty.
AB - Background: During the COVID-19 pandemic, there have been concerns regarding potential bias in pulse oximetry measurements for people with high levels of skin pigmentation. We systematically reviewed the effects of skin pigmentation on the accuracy of oxygen saturation measurement by pulse oximetry (SpO2) compared with the gold standard SaO2 measured by CO-oximetry. Methods: We searched Ovid MEDLINE, Ovid Embase, EBSCO CINAHL, ClinicalTrials.gov, and WHO International Clinical Trials Registry Platform (up to December 2021) for studies with SpO2–SaO2 comparisons and measuring the impact of skin pigmentation or ethnicity on pulse oximetry accuracy. We performed meta-analyses for mean bias (the primary outcome in this review) and its standard deviations (SDs) across studies included for each subgroup of skin pigmentation and ethnicity and used these pooled mean biases and SDs to calculate accuracy root-mean-square (Arms) and 95% limits of agreement. The review was registered with the Open Science Framework (https://osf.io/gm7ty). Results: We included 32 studies (6505 participants): 15 measured skin pigmentation and 22 referred to ethnicity. Compared with standard SaO2 measurement, pulse oximetry probably overestimates oxygen saturation in people with the high level of skin pigmentation (pooled mean bias 1.11%; 95% confidence interval 0.29 to 1.93%) and people described as Black/African American (1.52%; 0.95 to 2.09%) (moderate- and low-certainty evidence). The bias of pulse oximetry measurements for people with other levels of skin pigmentation or those from other ethnic groups is either more uncertain or suggests no overestimation. Whilst the extent of mean bias is small or negligible for all subgroups evaluated, the associated imprecision is unacceptably large (pooled SDs > 1%). When the extent of measurement bias and precision is considered jointly, pulse oximetry measurements for all the subgroups appear acceptably accurate (with Arms < 4%). Conclusions: Pulse oximetry may overestimate oxygen saturation in people with high levels of skin pigmentation and people whose ethnicity is reported as Black/African American, compared with SaO2. The extent of overestimation may be small in hospital settings but unknown in community settings. Review protocol registration: https://osf.io/gm7ty.
KW - Arterial blood oxygen saturation
KW - Ethnicity
KW - Measurement bias
KW - Pulse oximetry
KW - Skin pigmentation
KW - Systematic review
UR - http://www.scopus.com/inward/record.url?scp=85136012623&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85136012623&partnerID=8YFLogxK
U2 - 10.1186/s12916-022-02452-8
DO - 10.1186/s12916-022-02452-8
M3 - Article (journal)
C2 - 35971142
AN - SCOPUS:85136012623
SN - 1741-7015
VL - 20
SP - 1
EP - 14
JO - BMC Medicine
JF - BMC Medicine
IS - 1
M1 - 267
ER -